High Power Circuitry

  • Stan M. BachJr.
  • Paul Monroe
Chapter
Part of the Developments in Cardiovascular Medicine book series (DICM, volume 188)

Abstract

THE ICD must have two high power circuits in order to function. First, charging circuitry must convert the 6 V typically available from the battery to the much higher voltage (usually up to 750 V) necessary for successful defibrillation. Secondly, after this high voltage energy is concentrated in the capacitor, output switching circuitry then transfers this energy to the heart.

Keywords

Solder Joint Fixed Duration Electronic Switch Capacitor Charge Metal Oxide Semiconductor Field Effect Transistor 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Schuder JC, Stoeckle H, West JA, et al. Transthoracic ventricular defibrillation in the dog with truncated and untruncated exponential stimuli. IEEE Trans Bio-Med Eng 1971;18:410–415.CrossRefGoogle Scholar
  2. 2.
    Schuder JC, Gold JH, Stoeckle H, et al. Transthoracic ventricular defibrillation in the 100 kg calf with untruncated and truncated exponential stimuli. IEEE Trans Bio-Med Eng 1980;27:37–43.CrossRefGoogle Scholar
  3. 3.
    Keimel J, inventor. Medtronic Inc., assignee. Apparatus for delivering single and multiple cardioversion and defibrillation pulses. US Patent No. 5,163,427. 1992 Nov 17.Google Scholar
  4. 4.
    Schuder JC, McDaniel WC, Stoeckle H. Transthoracic defibrillation of 100 kg calves with bidirectional truncated exponential shocks. Trans Am Soc Artif Intern Organs 1984;30:520–525.PubMedGoogle Scholar
  5. 5.
    Bocchi DE, Laackman JT, Bach Jr SM, inventors. Cardiac Pacemakers Inc., assignee. Implantable N-phasic defibrillator output bridge circuit. US Patent No. 4,998,531. 1991 March 12.Google Scholar
  6. 6.
    Bach Jr SM, inventor. Cardiac Pacemakers Inc., assignee. Biphasic pulse generator for an implantable defibrillator. US Patent No. 4,850,357. 1989 JulyGoogle Scholar
  7. 7.
    de Coriolis PE, Batty Jr JR, Shook BJ, inventors. Telectronics Pacing Systems Inc., assignee. Method and apparatus for applying asymmetric biphasic truncated exponential countershocks. US Patent No. 5,083,562. 1992 Jan 28.Google Scholar
  8. 8.
    Winstrom WL, inventor. Intermedics Inc., assignee. Apparatus for generating multiphasic defibrillation pulse waveform. US Patent No. 4,800,883. 1989 Jan 31.Google Scholar
  9. 9.
    Pless B, Ryan JG, Culp JM, inventors. Ventritex, assignee. System configuration for combined defibrillator/pacer. US Patent No. 5,111,816. 1992 May 12.Google Scholar
  10. 10.
    Nunain SO, Roelke M, Trouton T, et al. Limitations and late complications of third generation automatic cardioverter-defibrillators. Circulation 1995;91:2204–2213.PubMedCrossRefGoogle Scholar
  11. 11.
    Cardiac Pacemakers Inc. Ventak P2 AICD summary of safety and effectiveness data. St. Paul, Minnesota, CPI/Guidant 1995.Google Scholar
  12. 12.
    Schuder JC, Stoeckle H, Gold JH, West JA, Keskar PY. Experimental ventricular defibrillation with an automatic and completely implanted system. Trans An Soc Artif Intern Organs 1970;16:207–212.Google Scholar