Troubleshooting Suspected ICD Malfunction

  • Mark E. Rosenthal
  • J. Todd Alderfer
  • Francis E. Marchlinski
Chapter
Part of the Developments in Cardiovascular Medicine book series (DICM, volume 188)

Abstract

THE ICD has made a dramatic impact on the management of patients with ventricular tachycardia (VT) and fibrillation (VF), markedly reducing the risk of sudden cardiac death in this high-risk group. In addition, the ICD has evolved from a relatively simple, non-programmable device to a complex unit with multiple options for tachycardia detection and therapy. As ICD technology has developed, a variety of clinical scenarios have arisen which force the electrophysiologist to differentiate appropriate from inappropriate device function. This chapter seeks to outline essentials of routine ICD clinical followup and then identify the most common types of clinical problems, other than infection, encountered during ICD followup. (Table 1) The differential diagnosis and plan for management of these problems will be discussed.

Keywords

Ventricular Tachycardia Automatic Gain Control Nonsustained Ventricular Tachycardia Defibrillation Threshold High Energy Discharge 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Corbelli R, McAllister H, Rashidi R, et al. Initial Beep-o-gram experience: Assessment of automatic implantable cardioverter-defibrillator sensing function. PACE 1986;12:665A. (abstract)Google Scholar
  2. 2.
    Chapman PD, Troup P. The automatic implantable cardioverter-defibrillator: Evaluating suspected inappropriate shocks. J Am Coll Cardiol 1986;7:1075–1081.PubMedCrossRefGoogle Scholar
  3. 3.
    Axtell KA, Akhtar M. Incidence of syncope prior to implantable cardioverter-defibrillator discharges. Circulation 1990;82:211A. (abstract)Google Scholar
  4. 4.
    Kou WH, Calkins HG, Lewis RR, et al. Incidence of loss of consciousness during automatic implantable cardioverter-defibrillator shocks. Ann Int Med 1991;115:942–945.PubMedGoogle Scholar
  5. 5.
    Grimm W, Flores BT, Marchlinski FE. Symptoms and electrocardiographically documented rhythm preceding spontaneous shocks in patients with implantable cardioverter-defibrillators. Am J Cardiol 1993;71:1415–1418.PubMedCrossRefGoogle Scholar
  6. 6.
    Marchlinski FE, Flores BT, Buxton AB, et al. The automatic implantable cardioverter-defibrillator: Efficacy, complications and device failures. Ann Int Med 1986;104:481–488.PubMedGoogle Scholar
  7. 7.
    Maloney J, Masterson M, Khoury D, et al. Clinical performance of the implantable cardioverter-defibrillator: electrocardiograpnic documentation or 101 spontaneous discharges. PACE 1991;14:280–285.PubMedCrossRefGoogle Scholar
  8. 8.
    Kelly PA, Mann DE, Damle RS, et al. Oversensing during ventricular pacing in patients with a third generation implantable cardioverter-defibrillator. J Am Coll Cardiol 1994;23:1531–1534.PubMedCrossRefGoogle Scholar
  9. 9.
    Newman D, Dorian P, Downar E, et al. Use of telemetry functions in the assessment of implanted antitachycardia device efficacy. Am J Cardiol 1992;70:606–621.CrossRefGoogle Scholar
  10. 10.
    Grimm W, Flores BF, Marchlinski FE. Complications of implantable cardioverter-defibrillator therapy: Followup of 241 patients. PACE 1993;16:218–222.PubMedCrossRefGoogle Scholar
  11. 11.
    Grimm W, Flores BT, Marchlinski FE. Shock occurrence and survival in 241 patients with implantable cardioverter-defibrillator therapy. Circulation 1993;87:1880–1888.PubMedCrossRefGoogle Scholar
  12. 12.
    Grimm W, Flores BT, Marchlinski FE. Electrocardiographically documented unnecessary, spontaneous shocks in 241 patients with implantable cardioverter-defibrillators. PACE 1992;15:1667–1673.PubMedCrossRefGoogle Scholar
  13. 13.
    Fogoros RN, Elson JJ, Bonnet CA. Actuarial incidence and pattern of occurrence of shocks following implantation of the automatic implantable cardioverter-defibrillator. PACE 1989;12:1465–1473.PubMedCrossRefGoogle Scholar
  14. 14.
    Reiter MJ, Mann DE. Sensing and tachyarrhythmia detection problems in implantable cardioverter defibrillators. J Cardiovasc Electrophysiol 1996;7:542–558.PubMedCrossRefGoogle Scholar
  15. 15.
    Hook BG, Callans DJ, Kleinman RB, et al. Implantable cardioverter-defibrillator therapy in the absence of significant symptoms: Rhythm diagnosis and management aided by stored electrograms. Circulation 1993;87:1897–1906.PubMedCrossRefGoogle Scholar
  16. 16.
    Edel TB, Maloney JD, Moore S, et al. Six-year clinical experience with the automatic implantable cardioverter-defibrillator. PACE 1991;14:1850–1854.PubMedCrossRefGoogle Scholar
  17. 17.
    Horton RP, Canby RC, Roman CA, et al. Diagnosis of ICD lead failure using continuous event marker recording. PACE 1995;18:1331–1334.PubMedCrossRefGoogle Scholar
  18. 18.
    Almeida HF, Buckingham TA. Inappropriate implantable cardioverter defibrillator shocks secondary to sensing lead failure: utility of stored electrograms. PACE 1993;16:407–411.PubMedCrossRefGoogle Scholar
  19. 19.
    Almassi GH, Olinger GN, Wetherbee JN, et al. Long term complications of implantable cardioverter defibrillator lead systems. Ann Thorac Surg 1993;55:888–892.PubMedCrossRefGoogle Scholar
  20. 20.
    Epstein AE, Shepard RB. Failure of one conduction in a nonthoracötomy implantable defibrillator lead causing inappropriate sensing and potentially ineffective shock delivery. PACE 1993;16:796–800.PubMedCrossRefGoogle Scholar
  21. 21.
    Bardy GH, Troutman C, Poole VE, et al. Clinical experience with a tiered therapy multiprogrammable antiarrhythmia device. Circulation 1992;85:1689–1698.PubMedCrossRefGoogle Scholar
  22. 22.
    Gottlieb C, Miller JM, Rosenthal ME, et al. Automatic implantable defibrillator discharge resulting from routine pacemaker programming. PACE 1988;11:336–338.PubMedCrossRefGoogle Scholar
  23. 23.
    Karson T, Grace K, Denes P. Stereo speaker silences automatic implantable cardioverter-defibrillator. Letters to Editor. N Engl J Med 1989;320:1628–1629.PubMedGoogle Scholar
  24. 24.
    Man KC, Davidson T, Langberg JT, et al. Interference from a hand held radiofrequency remote control causing discharge of an implantable defibrillator. PACE 1993;16:1756–1758.PubMedCrossRefGoogle Scholar
  25. 25.
    Madrid AH, Moro C, Martin J, et al. Interference of the implantable cardioverter defibrillator caused by slot machines. PACE 1996;19:675. (abstract)Google Scholar
  26. 26.
    Bárbaro V, Bartolini P, Donato A, et al. Do European GSM mobile cellular phones pose a potential risk to pacemaker patients? PACE 1995;18:1218–1224.PubMedCrossRefGoogle Scholar
  27. 27.
    Naegeli B, Osswald S, Deola M, et al. Intermittent pacemaker dysfunction caused by digital mobile telephones. J Am Coll Cardiol 1996;27:1471–1477.PubMedCrossRefGoogle Scholar
  28. 28.
    Fetter J, Ivan V, Benditt DG, et al. Digital cellular telephones do not necessarily interfere with implantable cardioverter defibrillator (ICD) operation. PACE 1996;19:676. (abstract)Google Scholar
  29. 29.
    Stanton MS, Grice S, Trusty J, et al. Safety of various cellular phone technologies with implantable cardioverter defibrillators. PACE 1996;19:583. (abstract)Google Scholar
  30. 30.
    Madrid AH, Moro C, Chadli, et al. Interferences between the automatic defibrillators and mobile phones. PACE 1996;19:676. (abstract)Google Scholar
  31. 31.
    McIvor ME. Environmental electromagnetic interference from electronic article surveillance devices; interactions with an ICD. PACE 1995;18:2229–2230.PubMedCrossRefGoogle Scholar
  32. 32.
    Ferrick KJ, Johnston D, Kim SG. Inadvertent AICD inactivation while playing bingo. Am Heart J 1991;121:206–207.PubMedCrossRefGoogle Scholar
  33. 33.
    Rosenthal ME, Paskman C. Noise generation during bradycardia pacing with the Ventritex Cadence/CPI Endotak ICD system: incidence and clinical significance. PACE 1996;19:677. (abstract)Google Scholar
  34. 34.
    Calkins H, Brinker J, Veltri EP, et al. Clinical interactions between pacemakers and automatic implantable cardioverter-defibrillators. J Am Coll Cardiol 1990;16:666–673.PubMedCrossRefGoogle Scholar
  35. 35.
    Kim SG, Furman S, Matos JA, et al. Automatic implantable cardioverter-defibrillator: Inadvertent discharges during permanent pacemaker magnet tests. PACE 1987;10:579–582.PubMedCrossRefGoogle Scholar
  36. 36.
    Callans DJ, Hook BG, Marchlinski FE. Paced beats following single nonsensed complexes in a “codependent” cardioverter defibrillator and bradycardia pacing system: Potential for VT induction. PACE 1991;14:1281–1287.PubMedCrossRefGoogle Scholar
  37. 37.
    Callans DJ, Hook BG, Kleinman RB, et al. Unique sensing errors in third generation implantable cardioverter-defibrillators. J Am Coll Cardiol 1993;22:1135–1140.PubMedCrossRefGoogle Scholar
  38. 38.
    Johnson NJ, Marchlinski FE. Arrhythmias induced by device antitachycardia therapy due to diagnostic specificity. J Am Coll Cardiology 1991;18:1418–1425.CrossRefGoogle Scholar
  39. 39.
    Hook BG, Marchlinski FE. Value of ventricular electrogram recordings in the diagnosis of arrhythmias precipitating electrical device shock therapy. J Am Coll Cardiol 1991;17:985–990.PubMedCrossRefGoogle Scholar
  40. 40.
    Callans DJ, Hook BG, Marchlinski FE. Use of bipolar recordings from patch-patch and rate sensing leads to distinguish ventricular tachycardia from supraventricular rhythms in patients with implantable cardioverter-defibrillators. PACE 1991;14:1917;-1922;.Google Scholar
  41. 41.
    Marchlinski FE, Gottlieb CD, Sarter B, et al. ICD data storage—value in arrhythmia management. PACE 1993;16:527–534.PubMedCrossRefGoogle Scholar
  42. 42.
    Brachman J, Sterns LD, Hilbel T, et al. Regularity of spontaneously occurring VT and its influence on stability criteria for detection. Circulation 1993;88:1353. (abstract)Google Scholar
  43. 43.
    Volosin KJ, Beauregard MA, Fabiszewski R. Spontaneous changes in ventricular tachycardia cycle length. J Am Coll Cardiol 1991;17:409–414.PubMedCrossRefGoogle Scholar
  44. 44.
    Geibel A, Zehender M, Brugada P. Changes in cycle length at the onset of sustained tachyarrhythmias—importance for antitachycardia pacing. Am Heart J 1988;115:588–592.PubMedCrossRefGoogle Scholar
  45. 45.
    Willems S, Borggrefe M, Shenasa M, et al. Radiofrequency catheter ablation of VT following implantation of an automatic cardioverter defibrillator. PACE 1993;16:1684–1692.PubMedCrossRefGoogle Scholar
  46. 46.
    Gottlieb CD, Horowitz LN. Potential interactions between antiarrhythmic medication and the automatic implantable cardioverter-defibrillator. PACE 1991;14:898–904.PubMedCrossRefGoogle Scholar
  47. 47.
    Hurwitz JL, Hook BG, Flores BT, et al. Importance of abortive shock capability with electrogram storage in cardioverter-defibrillator devices. J Am Coll Cardiol 1993;21:895–900.PubMedCrossRefGoogle Scholar
  48. 48.
    Swerdlow CD, Ahern T, Chen P, et al. Underdetection of VT by algorithms to enhance specificity in a tiered-therapy cardioverter-defibrillator. J Am Coll Cardiol 1994;24:416–424.PubMedCrossRefGoogle Scholar
  49. 49.
    Reiter Michael J, Div. Cardiology, Univ. of Colorado Health Sciences Center, Denver, CO 80262. Phone: 303-2/0-5173. Personal communication.Google Scholar
  50. 50.
    Miller JM, Hsia HH. Management of the patient with frequent discharges from implantable cardioverter devices. J Cardiovasc Electrophysiol 1996;7:278–285.PubMedCrossRefGoogle Scholar
  51. 51.
    Glikson M, Hammill SC, Hayes DL, et al. The importance of noise reversion in pacemaker-ICD interactions. PACE 1996;19:737. (abstract)CrossRefGoogle Scholar
  52. 52.
    Stambler BS, Wood MA, Damiano RJ, et al. Sensing/pacing lead complications with a newer generation implantable cardioverter-defibrillator: Worldwide experience from the Guardian ATP 4210 clinical trial. J Am Coll Cardiol 1994;23:123–132.PubMedCrossRefGoogle Scholar
  53. 53.
    Magney JE, Flynn DM, Parsons JA, et al. Anatomical mechanisms explaining damage to pacemaker leads, defibrillator leads, and failure of central venous catheters adjacent to the sternoclavicular joint. PACE 1993;16:445–457.PubMedCrossRefGoogle Scholar
  54. 54.
    Bardy GH, Ivey TD, Stewart R, et al. Failure of the automatic implantable defibrillator to detect ventricular fibrillation. Am J Cardiol 1986;58:1107–1108.PubMedCrossRefGoogle Scholar
  55. 55.
    Jung W, Manz M, Moosdorf R, et al. Failure of an implantable cardioverter-defibrillator to redetect ventricular fibrillation in patients with a nonthoraco-tomy lead system. Circulation 1992;86:1217–1222.PubMedCrossRefGoogle Scholar
  56. 56.
    Mann DE, Kelly PA, Damle RS, et al. Undersensing during ventricular tachyarrhythmias in a third generation implantable cardioverter-defibrillator. Diagnosis using stored electrograms and correction with programming. PACE 1994;17:1525–1530.PubMedCrossRefGoogle Scholar
  57. 57.
    Berul CI, Callans DJ, Schwartzman DS, et al. Comparison of initial detection and redetection of ventricular fibrillation in a transvenous defibrillator system with automatic gain control. J Am Coll Cardiol 1995;25:431–436.PubMedCrossRefGoogle Scholar
  58. 58.
    Ellis JR, Martin DT, Venditti FJ. Appropriate sensing of ventricular fibrillation after failed shocks in a transvenous cardioverter-defibrillator system. Circulation 1994;90:1820–1825.PubMedCrossRefGoogle Scholar
  59. 59.
    Singer I, Guarneri T, Kupersmith J. Implanted automatic defibrillators: Effects of drugs and pacemakers. PACE 1988;11:2250–2262.PubMedCrossRefGoogle Scholar
  60. 60.
    Kim SG, Furman S, Waspe LE, et al. Unipolar pacer artifacts induced failure of an automatic implantable cardioverter/defibrillator to detect ventricular fibrillation. Am J Cardiol 1986;57:880–889.PubMedCrossRefGoogle Scholar
  61. 61.
    Swerdlow CD, Chen PS, Kass RM, et al. Discrimination of ventricular tachycardia from sinus tachycardia and atrial fibrillation in a tiered therapy cardioverter-defibrillator. J Am Coll Cardiol 1994;23:1342–1355.PubMedCrossRefGoogle Scholar
  62. 62.
    Lehmann MH, Thomas A, Nabih M, et al. Sudden death in recipients of first-generation implantable cardioverter-defibrillators: analysis of terminal events. J Interventional Cardiol 1994;7:487–503.CrossRefGoogle Scholar
  63. 63.
    Saksena S, Polzobutt-Johanos M, Castle LW, et al. Long-term multicenter experience with a second generation implantable pacemaker-defibrillator in patients with malignant ventricular arrhythmias. J Am Coll Cardiol 1992;19:490–499.PubMedCrossRefGoogle Scholar
  64. 64.
    Jones GK, Bardy GH, Kudenchuk PJ, et al. Mechanical complications following implantation or multiple lead nonthoracotomy defibrillator systems: implications for management and future system design. Am Heart J 1995;10:327–333.CrossRefGoogle Scholar
  65. 65.
    Frame R, Brodman R, Furman S, et al. Long-term stability of defibrillation thresholds with intrapericardial defibrillator patches. PACE 1993;16:208–212.PubMedCrossRefGoogle Scholar
  66. 66.
    Venditti FJ, Martin DT, Vassolas G, et al. Rise in chronic defibrillation thresholds in nonthoracotomy implantable defibrillator. Circulation 1994;89:216–223.PubMedCrossRefGoogle Scholar
  67. 67.
    Guarneri T, Levine JH, Veltri EP, et al. Success of chronic defibrillation and the role of antiarrhythmic drugs with the automatic implantable cardioverter defibrillator. Am J Cardiol 1987;60:1061–1064.CrossRefGoogle Scholar
  68. 68.
    Haberman RJ, Veltri EP, Mower MM. The effect of amiodarone on defibrillation threshold. J Electrophysiol 1988;5:415–423.Google Scholar
  69. 69.
    Marinchak RA, Friehling T, Kline RA, et al. Effect of antiarrhythmic drugs on defibrillation threshold: Case report of an adverse effect of mexiletine and review of the literature. PACE 1988;11:7–12.PubMedCrossRefGoogle Scholar
  70. 70.
    Pinski SL, Vanerio G, Castle LW, et al. Patients with a high defibrillation threshold: Clinical characteristics, management and outcome. Am Heart J 1991;122:89–95.PubMedCrossRefGoogle Scholar
  71. 71.
    Epstein AE, Ellenbogen KA, Kirk KA, et al. Clinical characteristics and outcome of patients with high defibrillation thresholds: a multicenter study. Circulation 1992;86:1200–1216.Google Scholar
  72. 72.
    Haddad L, Padula LE, Moreau M, et al. Troubleshooting implantable cardioverter defibrillator system malfunctions: the role of impedance measurements. PACE 1994;17:1456–1461.PubMedCrossRefGoogle Scholar
  73. 73.
    Hayes DL, Vlietstra RE. Pacemaker malfunction. Ann Int Med 1993;119:828–835.PubMedGoogle Scholar
  74. 74.
    Marchlinski FE, Buxton AE, Flores B. Automatic Implantable Cardioverter Defibrillator: Followup and complications. In: Samet and El-Sheriff (eds:) Cardiac Pacing and Electrophysiology, 3rd Ed., Grune and Stratton 1990; Chapter 45, pp 743–758.Google Scholar
  75. 75.
    Marchlinski FE, Kleiman RB, Hook BG, et al. Advances in implantable cardioverter defibrillator therapy. In: Josephson ME, Wellens HJJ (eds): Tachycardias: Mechanisms and Management, Mt. Kisco, NY, Futura Publishing Co., Inc., 1993: pp405–420.Google Scholar

Copyright information

© Springer Science+Business Media New York 1996

Authors and Affiliations

  • Mark E. Rosenthal
  • J. Todd Alderfer
  • Francis E. Marchlinski

There are no affiliations available

Personalised recommendations